Agreement of the Point of Care Test (POCT) Boditech iCHROMA™ COVID-19 IgG Antibody Assay with the Abbott Architect SARS-CoV-2 IgG Antibody Assay: An Advanced Study Approach

Bass, M. and Bolodeoku, J. and Stevenson, E. and Anyaeche, C. and Kim, T. K. and Retnasingham, V. (2022) Agreement of the Point of Care Test (POCT) Boditech iCHROMA™ COVID-19 IgG Antibody Assay with the Abbott Architect SARS-CoV-2 IgG Antibody Assay: An Advanced Study Approach. In: Issues and Developments in Medicine and Medical Research Vol. 3. B P International, pp. 26-32. ISBN 978-93-5547-454-4

Full text not available from this repository.

Abstract

Background: Molecular tests are generally not suited for public health screening, antibody tests are more suited and the more portable, the better. The Boditech iCHROMA™ point of care immunoassay analyser has a Covid-19 antibody assay for the qualitative determination of IgG/IgM antibodies in human whole blood/serum/plasma. It is helpful as an aid in the screening of early mild asymptomatic or acute patients for identifying Covid-19 infection with high sensitivity.

Objectives: To determine the clinical agreement (sensitivity and specificity) of the Boditech iCHROMA™ Covid-19 antibody assay and the Abbott Architect SARS-CoV-2 IgG assay for the qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma.

Results: Of the 50 plasma samples, 20 (40%) of the plasma samples assayed were reported as positive by both the Abbott Architect SARS-CoV-2 IgG assay and the Boditech iCHROMA™ Covid-19 IgG antibody assay. The remaining 30 (60%) of the plasma samples were reported as negative by Abbott Architect SARS-CoV-2 IgG antibody assay. However, the Boditech iCHROMA™ Covid-19 IgG antibody assay, reported only 27 of the 30 (90%) samples as negative.

Conclusion: There was an overall agreement of 95%, with a sensitivity of 100% and a specificity of 90% of the Boditech iCHROMA™ Covid-19 IgG antibody assay and the Abbott Architect SARS-CoV-2 IgG assay.